Wireless Diagnostic Calibrator – PIE 901
Milliamp, voltage and pressure calibrator$1,765.00 – $1,850.00
New PIE 901 milliamp, voltage & pressure calibrator – Find Hidden Loop Problems!
- Calibrate Milliamp Instruments
Source 0.000 to 24.000 mA & -25.00 to 125.00% of 4-20 mA
Read 0.000 to 24.000 mA & -25.00 to 125.00% of 4-20 mA
Simulate 2-Wire Transmitter, Power & Read Transmitters
Accurate to ±(0.01% of Reading + 0.002 mA) - Calibrate Voltage Instruments
Source 0.000 to 24.000 VDC
Read -60.000 to 60.000 VDC
Accurate to ± (0.02% of Reading + 0.01% Full Scale) - Calibrate pressure instruments with optional pressure modules
Dual display of milliamp and pressure for calibration of I/P or P/I transducers and pressure gauges
Choose from a wide selection of pressure modules along with hand pumps and tubing kits - Blue tooth functionality (for use with future PIE wireless pressure modules)
- Large Color Display with color coded menus for simplified setup
- Locate Loop Current Leakages
Automatic indication of Loop Current and Leakage Current. Measure ground current leakage from faulty wiring, flooded conduit and corrosion bridges. Leakage typically appears as a Zero Shift and helps you decide if the transmitter needs calibration or if there is a problem in the loop wiring. - Faster Calibrations
Three output EZ-CHECK switch
2 speed EZ-DIAL
Stepping & Ramping - Includes Rubber Boot, Deluxe Hands Free Carrying Case, USBC Charger, 4 NiMh Rechargeable Batteries & Test Leads
- Optional Accessories
Magnetic hanging strap (020-0236)
Find Hidden Loop Problems with Ground Leak Detection
If you find a loop where the transmitter is calibrated correctly but all the readings elsewhere in the loop have a fixed offset this is due to a Zero Shift. This zero shift is typically caused by some current in the loop bypassing the transmitter. This might be caused by ground faults, moisture or corrosion.
Turn on Ground Leak Detection and use the PIE 901 to power up the loop. Any current that isn’t controlled by the transmitter or other current control element will be indicated as leakage on the calibrator’s display.
Typical problem found with Loop Diagnostics
Here is a loop where a technician has just recalibrated the transmitter but the control room still sees a problem. The problem started just after a rainstorm.
Using a PIE 901 to power the loop the technician detects a leakage of 0.51 mA – the same offset seen in the control room. He walks the loop and opens a junction box releasing a stream of water. The loop is again in control. Clink on the link Video of Loop Diagnostics to see a video demonstration of Loop Diagnostics in action.
Weight | 2 lbs |
---|---|
Dimensions | 7.75 × 5.25 × 4.5 in |
Accessory Package |
Standard Package ,Basic Package |
Related products

Wireless Diagnostic Pressure Module
$2,199.004-20 mA Loop Calibrator – PIE 334
Industry standard replaces Altek 334 $999.00
Multifunction Process Calibrator – PIE 820
6 functions in 1 calibrator $2,804.00 – $2,984.00
Diagnostic RTD and mA Calibrator – PIE 311Plus
2 units in 1, with patented RTD wire detection $2,454.00 – $2,634.00
mA Loop Diagnostic Calibrator – PIE 334Plus
4-20 milliamp loop calibrator with current leak detection $1,070.00 – $1,140.00
Diagnostic Temperature and mA Calibrator – PIE 525Plus
RTD, thermocouple and loop calibrator $2,894.00 – $3,074.00
10-50 mA Multifunction Process Calibrator – PIE 850
Loop diagnostics $4,954.00 – $5,134.00
Reviews
There are no reviews yet.